Waveform Amplifier
Instruments For Testing Your Innovations
Electrical Specifications
Output Voltage Range
TS250-0 -10V to +10V
TS250-1 -20V to +20V
TS250-2 -30V to +30V
TS250-3 -40V to +40V
TS250-4 -6V to +15V
TS250-5 -6V to +30V
TS250-6 -6V to +45V
TS250-7 -6V to +65V
Output Current (> 10Hz, VDC or Vpp = 0.75 Max, Resistive Load)
5.0ADC, -6.0A to +6.0A peak
3.1ADC, -4.4A to +4.4A peak
2.1ADC, -3.0A to +3.0A peak
1.7ADC, -2.5A to +2.5A peak
4.0ADC, -5.0A to +5.0A peak
2.1ADC, -3.0A to +3.0A peak
1.7ADC, -2.5A to +2.5A peak
2.1ADC, -2.5A to +2.5A peak
Maximum Frequency (Note 1)
3.5Vpp
10Vpp
20Vpp
40Vpp
80Vpp
Output Noise (Note 2)
RMS
Peak-to-Peak
Output Slew Rate
Input Impedance
Hi-Z setting
50Ω setting
Min Frequency
DC-Couple setting
AC-Couple and Hi-Z setting, -3dB
AC-Couple and 50Ω setting, -3dB
Voltage Gain
0dB setting
20dB setting
Current Monitor Scale (Note 3)
Current Monitor Accuracy (Max Current)
LCD Volt Meter Accuracy (Note 4)
LCD Ammeter Meter Accuracy (Note 4)
AC Power Supply Input
Operating Temperature Range
Storage Temperature Range
Operating Humidity Range (No Condensation)
Weight
Dimension
1MHz
450kHz
225kHz
112kHz
35kHz
1mVrms
6mVpp
16V/us
1kΩ
50Ω
DC
1Hz
50Hz
0dB (gain of 1V/V)
20dB (gain of 10V/V)
100mV/A
+/-1%
+/-1%
+/-1%
100VAC to 240VAC
0C to +40C
-20C to +60C
20% to 80%
4.0kg
11.25 x 10.5 x 5.5 Inches
1. The maximum frequency is calculated by fMAX = 4500/Vpp, in kHz. For example, the maximum frequency for a 20Vpp (-10V to +10V sinewave) is 225kHz.
2. 0dB gain, 50Ω setting, current monitor output is connected to an oscilloscope.
3. 100mV is equal to 1A output current when terminated with 50 ohm. Output voltage is twice if not terminated.
4. LCD meter reading is the average voltage or current. For example, the average voltage of a sinewave is 0V.
5. Specifications are subject to change without notice.